16 Special Functions
16.2 Batch Schedule
225
GC-2010 Plus
16.2.5 Parameter list
START VIAL NO.
Range:
1
Default: 0
FINAL VIAL NO.
Range:
1
Default: 0
1: With no auto sampler
Short rack ... 1−6
Long rack ... 1−12 (1−15 allowable, with an additional rack)
With auto sampler 1−150(maximum; for 1.5 mL vial)
1−96 (maximum; for 4 mL vial)
When the auto sampler is used, the possible depends on the vial rack type and the number
of racks.
The default, “0”, indicates that the setting is not given.
SAMPLE
Inject sample
The specified sample vial No. is analysed.
Without sample
Vials are not used for samples. Instead samples are introduced from a gas sampler with a
valve.
GC FILE NO.
Range: FILE 0−9, Default: FILE0
Set the file No. of the GC method to be used.
CHANNEL NO.
Range: Ch1/Ch2, Default: Ch 1
Set the channel No. connecting the GC to the Chromatopac.
DATA PROCESSOR FILE NO.
Range: FILE 0−9, Default: FILE0
Specify a date processor (Chromatopac) date integration file.
NUMBER OF INJECTION
Range: 0−99, Default: 0
Set the number of times to inject each sample.
SAMPLE TYPE
Unknown Select this item to analyze an unknown sample.
Analyze a standard and create a calibration curve before quantifying an unknown
sample.
Standard Select this item to analyze a standard for creating a calibration curve.
SLOPE TEST
Range: On/Off, Default: Off
When “On” is selected, the slope test is performed before running the batch schedule.
CLEAN UP
Range: On/Off, Default: Off
When “On” is selected, clean up is performed before running the batch schedule.