15 Optional Devices
15.4 Setting the CRG Parameters
212
GC-2010 Plus
15.4.2 Parameter list
COLUMN OVEN
VALVE Selection: On/Off, Default: Off
CRG Selection: Use/Not Use, Default: Not Use
When the CRG is connected to the column oven, the column oven temperature
control range is extended to -99.0 − +450.0 °C.
When using the CRG, set “VALVE” to “On” and “CRG” to “Use”.
When using valve control only, set “VALVE” to “On” and set “CRG” to “Not Use”.
INJ 2
VALVE Selection: On/Off, Default: Off
CRG Selection: Use/ Not Use, Default: Not Use
When the PTV is installed on INJ2 and the CRG is connected, the injection port
temperature control range is extended to -99.0 − +450.0 °C.
When using the CRG, set “VALVE” to “On” and “CRG” to “Use”.
When using valve control only, set “VALVE” to “On” and set “CRG” to “Not Use”.
15.4.3 PF menu
PF menu Description
Reference
section
Next
Toggles among the AOC parameters → AUX temperature → AUX APC →
CRG screens, inthat order.
――