16 Special Functions
16.2 Batch Schedule
228
GC-2010 Plus
16.2.9 Batch processing Setup example
Example 1: For single AOC
Sample vials No. 1 to 3 are analyzed. The GC method file No. 2 is used. The
Chromatopac data processing file No. 0 is set to Ch 1.
A standard is injected once.
Clean up is performed, but the slope test is not performed.
Screen display
1. Start vial No.: 1
Final vial No.: 3
2. GC file No.: FILE2
3. Data processing file No.:
Ch 1, FILE0
4. Number of injections: 1
Sample type: Standard
5. Slope test: Off
Clean up: On
6. The batch is set up.