2 Installation
2.3 Installation site
14
GC-2010 Plus
Q Installation clearances
Fig. 2.3.1  
2.3.6 Rear clearance
Hot air is vented at the back of the unit when the column oven cools. Consider the following
during installation:
Do not place any flammable materials behind the unit.
Allow a clearance of 50 cm or more between the back cover and the wall.
(*)
Reserve extra space for maintenance and inspection behind the unit.
() When the optional exhaust duct (221-47748-41) is used, the clearance from the wall can be a
minimum of 25 cm.
2.3.7 Left clearance
Allow a clearance of 10 cm or more on the left side. The vent on the left side is important for
properly cooling down the outer wall of the column oven and improving the cooling efficiency
of the column oven.
WARNING
Hot air is exhausted from the back vent. Do not place flammable
matelials near the exhaust.
GC-2010 Plus
Minimum 500mm
Minimum
100mm
Depth
530mm
Hight
440mm
Width 515mm

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2010 Plus and is the answer not in the manual?

Shimadzu GC-2010 Plus Specifications

General IconGeneral
TypeGas Chromatograph
ModelGC-2010 Plus
ManufacturerShimadzu
Maximum Temperature Ramp Rate120 °C/min
Detector OptionsFID, TCD, ECD, FPD, NPD, MS
Electronic Flow Control (EFC)Yes
Data Sampling RateUp to 250 Hz
Column Oven Temperature RangeAmbient +5°C to 450°C
Maximum Oven Heating Rate120 °C/min
Injection PortSplit/Splitless, PTV
DisplayLCD

Related product manuals