16 Special Functions
16.3 Time Program
233
GC-2010 Plus
Event91 and Event92: Switches a-contact and b-contact of each relay on the CPU PCB
(standard).
Event of PRG: Switches a-contact and b-contact of each relay on the PRG PCB (option).
Event of On/Off point: Turns the relay contact on the PRG PCB (option) on/off.
AC On/Off: Turns the power supplied to the AC connector on the power controller PCB on/off.
Blower, Cooling Fan, CRG INJ and CRG Column: Turns the power of each option on/off.
2: Current range
16.3.6 Editing a time program
To change the contents of an existing time program, move the cursor using the [ △ ] and [ ▽ ]
keys to the program line to be edited changed and press the [ENTER] key.
To change only the time, enter the new time from the edit screen and press the [ENTER]
key. Then press [Finish] (PF menu).
To change the other parameters, move the cursor and make the changes from the edit
screen, and then press [Finish] (PF menu).
To cancel the edits and restore the original time program, press [Cancel] (PF menu) before
pressing [Finish] (PF menu).
Detector type Current
FTD
0.00−10.00 (pA)
ECD
0.00−2.00 (nA)
TCD
0−100 (mA)