EasyManuals Logo

Shimadzu GC-2010 Plus User Manual

Shimadzu GC-2010 Plus
352 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #253 background imageLoading...
Page #253 background image
16 Special Functions
16.3 Time Program
233
GC-2010 Plus
Event91 and Event92: Switches a-contact and b-contact of each relay on the CPU PCB
(standard).
Event of PRG: Switches a-contact and b-contact of each relay on the PRG PCB (option).
Event of On/Off point: Turns the relay contact on the PRG PCB (option) on/off.
AC On/Off: Turns the power supplied to the AC connector on the power controller PCB on/off.
Blower, Cooling Fan, CRG INJ and CRG Column: Turns the power of each option on/off.
2: Current range
16.3.6 Editing a time program
To change the contents of an existing time program, move the cursor using the [ ] and [ ]
keys to the program line to be edited changed and press the [ENTER] key.
To change only the time, enter the new time from the edit screen and press the [ENTER]
key. Then press [Finish] (PF menu).
To change the other parameters, move the cursor and make the changes from the edit
screen, and then press [Finish] (PF menu).
To cancel the edits and restore the original time program, press [Cancel] (PF menu) before
pressing [Finish] (PF menu).
Detector type Current
FTD
0.0010.00 pA
ECD
0.002.00 nA
TCD
0100 mA

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2010 Plus and is the answer not in the manual?

Shimadzu GC-2010 Plus Specifications

General IconGeneral
TypeGas Chromatograph
ModelGC-2010 Plus
ManufacturerShimadzu
Maximum Temperature Ramp Rate120 °C/min
Detector OptionsFID, TCD, ECD, FPD, NPD, MS
Electronic Flow Control (EFC)Yes
Data Sampling RateUp to 250 Hz
Column Oven Temperature RangeAmbient +5°C to 450°C
Maximum Oven Heating Rate120 °C/min
Injection PortSplit/Splitless, PTV
DisplayLCD