EasyManuals Logo

Shimadzu GC-2010 Plus User Manual

Shimadzu GC-2010 Plus
352 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #319 background imageLoading...
Page #319 background image
18 Maintenance and Inspection
 
18.7
18.
299
GC-2010 Plus
18.7
Inspection and Maintenance: Capillary Column
NOTE For the details on removing the capillary column, refer to “4 Installing the column”.
18.7.1 Inspection/maintenance cycle
In the following situations, inspect and/or maintain the capillary column.
When using a new column or a column which has not been used for a long time
When ghost peaks are detected
When the baseline is unstable
When the baseline noise is high
18.7.2 Inspection/maintenance
Q Troubleshooting
Check whether the baseline is unstable or ghost peaks are present.
A contaminated column can cause an unstable baseline or ghost peaks. If these problems
occur, the column should be conditioned.
Q Preparing the gas chromatograph main body
If the system is operating, press the [SYSTEM] key and select [Stop GC] (PF menu). The
temperature of the injection port, the column oven and detector automatically decrease.
Press the [MONIT] key, and ensure that the temperature of the injection port, the detector
and the column are below 50 °C
Press the [FLOW] key, and select [Off] (PF menu). The inspection/maintenance can then
begin.
NOTE For the details on handling the capillary column, refer to “4 Installing the column”.
Q Conditioning the capillary column
Condition the capillary column using the following procedure.
(1) Setting the carrier gas
To condition the column, use the an aleysis flow rate. However, reduce the split ratio to
conserve carrier gas.
(2) Setting the column oven temperature
Set the column oven to a temperature approximately 30 °C higher than the column
operating temperature during analyses.
Do not exceed the maximum column temperature limit.
(3) Setting the injection port temperature and the detector temperature
Use the same temperatures normally set for analyses.
Set the detector to a temperature approximately 30 °C higher than the column operating
temperature.
WARNING
Risk of burns. Do not perform maintenance until the temperature of
the injection port, detecor, and oven have dropped below 50 °C

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2010 Plus and is the answer not in the manual?

Shimadzu GC-2010 Plus Specifications

General IconGeneral
TypeGas Chromatograph
ModelGC-2010 Plus
ManufacturerShimadzu
Maximum Temperature Ramp Rate120 °C/min
Detector OptionsFID, TCD, ECD, FPD, NPD, MS
Electronic Flow Control (EFC)Yes
Data Sampling RateUp to 250 Hz
Column Oven Temperature RangeAmbient +5°C to 450°C
Maximum Oven Heating Rate120 °C/min
Injection PortSplit/Splitless, PTV
DisplayLCD

Related product manuals