16 Special Functions
16.1 Time Scheduler
217
GC-2010 Plus
16.1.8 Parameters available in the time scheduler
Parameter Description
Relay
1
Detector
Zero Adj Turns zero adjustment for each detector on/off.
Flame Turns ignition on/off.
FTD
Turns voltage feedback on/off to make current constant for
each detector.
Detector controller Turns control of each detector on/off during analysis program.
Flow
controller
Splitter Ctrl Opens/closes splitter.
Air(value) Opens/closes air solenoid valve.
Flow controller On/Off Turns control of flow controller on/off.
High press inj mode Turns high pressure injection mode on/off.
Gas saver Turns gas saver split rate function on/off for each injection port.
Temperature Turns all heated zones on/off.
Stop
NUMBER
OF
TIMES
OF RUN
Time
scheduler
stop
Specifies the number of times the program can execute before
the schedule stops or switches to another file. If greater than
one, the choices are “Continuous” or “Pause at end of repeat”.
For an AOC, select “Pause at end of repeat”. When
“Continuous” is chosen,the program executes at set intervals
continuously at set intervals.
If another program (such as a temperature program) is running
when time schedule execution time expires, priority is given to
the running program. When the current program finishes, the
time schedule stops or file is changed to another file.
Schedule
0-9 Load
File load Selects the next file when switching to another file.
Program
start
Program
Starts gas chromatograph after reaching ready state, or starts it
immediately.
Clean Up
Starts clean up program after reaching ready state, or starts it
immediately.
Pre-Run program
Starts Pre-Run program after reaching ready state, or starts it
immediately.
GC Start/Stop Sets start/stop of gas chromatograph.
Other
Set the parameter to input event No.
(Refer to “17.4 Event No.”.)