15.1
15.
197
GC-2010 Plus
15 Optional Devices
15.1Auto Injector Parameters
From the gas chromatograph, specify the parameters of the Shimadzu AOC-20i Auto Injector and the
AOC-20s Auto Sampler Carousel, which automatically inject liquid samples into the gas chromatograph.
For detailed setting procedures, refer to the User's Manual for AOC-20 and the User’s Manual for
AOC-20 Dual Injection System.
15.1.1 Screen description
After installing the auto injector on the gas chromatograph, access the line configuration
screen from the [SET] key, and then select AOC1 for the analytical line (select AOC2 in addi-
tion, for the analytical line for dual injection systems.
)
Press the [OPTION] key again or press [Next] (PF menu) several times on the [OPTION] key
screen to display the AOC parameter screen shown in Fig. 15.1.1. The screen toggles
among the AOC parameters
→ AUX temperature → AUX APC → CRG screens.
Q AOC STATUS
The AOC status can be monitored.
Fig. 15.1.1 AOC setup screen
Screen display Description
Inactive AOC is not operating.
Taking vial A vial is being picked up from the tray.
Preparation AOC is operating (before injection).
Rinse AOC is operating (after injection).
Putting vial A vial is being returned to the tray.
Wait AOC is waiting for analysis.
The AOC status is displayed.