7 Starting and Stopping the GC [SYSTEM]
 
7.3
7.
78
GC-2010 Plus
7.3Specifying Start Procedures
7.3.1 Screen description
Select [Start Seq] (PF menu) from the [SYSTEM] key main screen to display the start
procedure setup screen shown in Fig. 7.3.1.
On this screen, set whether the system automatically starts (Auto Start) when the power is
next turned on, or whether the system does not start until [Start GC] (PF menu) is pressed
(Manual Start). Alternatively, only carrier gas flow is turned on the next time the power is
turned on. To start the system, select [Start GC] (PF menu) as for a manual start. This is
known as a semi-auto start. Finally, an analysis file can be set to begin the next time the
power is turned on or the GC is restarted. Any settings changed on this screen are reflected
in the [SYSTEM] key main screen.
Fig. 7.3.1   Setting the start procedures for the next GC restart

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2010 Plus and is the answer not in the manual?

Shimadzu GC-2010 Plus Specifications

General IconGeneral
TypeGas Chromatograph
ModelGC-2010 Plus
ManufacturerShimadzu
Maximum Temperature Ramp Rate120 °C/min
Detector OptionsFID, TCD, ECD, FPD, NPD, MS
Electronic Flow Control (EFC)Yes
Data Sampling RateUp to 250 Hz
Column Oven Temperature RangeAmbient +5°C to 450°C
Maximum Oven Heating Rate120 °C/min
Injection PortSplit/Splitless, PTV
DisplayLCD

Related product manuals