16 Special Functions
16.1 Time Scheduler
216
GC-2010 Plus
16.1.6 Creating a new time schedule
Select [New] (PF menu) from the time schedule screen to display the Time Schedule screen
shown in Fig. 16.1.5.
Set an item using the [] and [] keys and [ENTER] key. When [Finish] (PF menu) appears
after you have set items, press [Finish] to validate the schedule. The set items will not be
updated or saved unless [Finish] is pressed.
In the upper half of the screen, the contents of the set schedule are displayed. In the lower
half of the screen, the schedule edit screen is displayed.
Up to 100 lines are available per schedule.
The schedule execution time range is from 00:00 to 23:59. The unit is 1 minute.
16.1.7 PF menu
Fig. 16.1.5   Creating a new schedule
PF menu Description
Reference
section
Finish Validates and saves the time schedule.
――
Prev Page Returns to previous setup screen.
――
Next Page
Moves to next item.
The cursor can also be moved to next item by pressing [ENTER] key.
――
Cancel Cancels the time schedule currently being created.
――

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2010 Plus and is the answer not in the manual?

Shimadzu GC-2010 Plus Specifications

General IconGeneral
TypeGas Chromatograph
ModelGC-2010 Plus
ManufacturerShimadzu
Maximum Temperature Ramp Rate120 °C/min
Detector OptionsFID, TCD, ECD, FPD, NPD, MS
Electronic Flow Control (EFC)Yes
Data Sampling RateUp to 250 Hz
Column Oven Temperature RangeAmbient +5°C to 450°C
Maximum Oven Heating Rate120 °C/min
Injection PortSplit/Splitless, PTV
DisplayLCD

Related product manuals