16 Special Functions
16.6 GC Configuration
254
GC-2010 Plus
16.6.10 Temperature offset
If the actual temperature of the various heated zones (as measured by a thermometer)
differs from the temperature displayed on the GC screen, set a temperature offset value to
compensate.
16.6.10.1 Screen description
Select “6. GC CONFIGURATION” from the [FUNC] key and select “8. TEMPERATURE
OFFSET” to display the temperature offset screen shown in Fig. 16.6.20.
16.6.10.2 Parameter list
COLUMN/INJ1/INJ2/DET1/DET2/AUX3/AUX4/AUX5
Range: -10.00−+10.00, Default: 0.00
The temperature of each heated zone can be offset, Uninstalled ports are not displayed.
For example, when “-1 °C” is obtained by subtracting the column oven temperature
measured by thermometer from the gas chromatograph value with an offset value of
“0”, input “-1” as the offset value.
Fig. 16.6.20 Offset set up screen