16 Special Functions
16.6 GC Configuration
254
GC-2010 Plus
16.6.10 Temperature offset
If the actual temperature of the various heated zones (as measured by a thermometer)
differs from the temperature displayed on the GC screen, set a temperature offset value to
compensate.
16.6.10.1 Screen description
Select “6. GC CONFIGURATION” from the [FUNC] key and select “8. TEMPERATURE
OFFSET” to display the temperature offset screen shown in Fig. 16.6.20.
16.6.10.2 Parameter list
COLUMN/INJ1/INJ2/DET1/DET2/AUX3/AUX4/AUX5
Range: -10.00+10.00, Default: 0.00
The temperature of each heated zone can be offset, Uninstalled ports are not displayed.
For example, when “-1 °C” is obtained by subtracting the column oven temperature
measured by thermometer from the gas chromatograph value with an offset value of
“0”, input “-1” as the offset value.
Fig. 16.6.20   Offset set up screen

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Shimadzu GC-2010 Plus and is the answer not in the manual?

Shimadzu GC-2010 Plus Specifications

General IconGeneral
TypeGas Chromatograph
ModelGC-2010 Plus
ManufacturerShimadzu
Maximum Temperature Ramp Rate120 °C/min
Detector OptionsFID, TCD, ECD, FPD, NPD, MS
Electronic Flow Control (EFC)Yes
Data Sampling RateUp to 250 Hz
Column Oven Temperature RangeAmbient +5°C to 450°C
Maximum Oven Heating Rate120 °C/min
Injection PortSplit/Splitless, PTV
DisplayLCD

Related product manuals