6 Analysis Flow Chart
6.1 Analysis Flow Chart
66
GC-2010 Plus
6.1.3 Analysis
Set the data processing Perform the required settings for the data processing unit, such as
unit specifying the processing parameters.
Check the baseline Press the [MONIT] key, and ensure that the baseline is stable.
Press [Zero Adj] (PF menu) to zero the detector output. When the
baseline is stable, you can start analysis.
Inject the sample Aspirate the sample in the syringe, inject it into the GC injection port,
and press [START] to analyze it.
For capillary column analysis, normally inject 1 µL or less liquid.
(1-2 µL available)
WARNING
Wear protective goggles when using a
syringe to inject samples.
The syringe plunger could be expelled
due to injection port back pressure.
Sample could get into the eyes.
By holding and supporting the plunger
from the side with your middle finger,
you can smoothly inject the sample and
keep the plunger in the syringe.
Do not bend the plunger when holding
the syringe in this position.
Hold here
(Example)
How to hold a syringe
when injection