1 Overview
1.2 Specifications
3
GC-2010 Plus
OCI/PTV-2010
Injection unit : On-column/PTV injection
Range of temperature : Room temperature + 5 °C to 450 °C (column oven temperature: 50 °C)
Program ramps : 7 ramps
Temperature setting : 0.1 °C increments
Program rate : Up to 250 °C/min
Heating speed : Within 3 minutes heating from 50 °C to 450 °C
Cooling speed : Approximately 8 minutes cooling from 450 °C to 50 °C (column
oven temperature: 50 °C)
Cooling : A special cooling fan is used.
1.2.4 Detector
Overheat protection : Programmable up to 470 °C
Time constant : 4 ms to 2 s selectable
Hydrogen flame ionization detector (FID)
Range of temperature : Up to 450 °C, 0.1 °C increments
Minimum detection : 1.5 pg C/s
Dynamic range : 10
7
Jet material : Fused quartz
Thermal conductivity detector (TCD)
Amplifier : With differential type filament protective circuit, tungsten
rhenium filament, dual-element configuration
Power : Constant current
Range of temperature : Up to 400 °C
Sensitivity : 20,000 mV ml/mg
Dynamic range : 10
5
Electron capture detector (ECD)
Radiation source :
63
Ni 370MBq
Range of temperature : Up to 350 °C
Minimum detection : 6 fg/s
Dynamic range : 10
4
Flame photometric detector (FPD)
Range of temperature : Up to 350 °C
Minimum detection : P: 55 fgP/s, S: 3 pgS/s
Dynamic range : P: 10
4
, S: 10
3
Flame thermionic detector (FTD)
Range of temperature : Up to 450 °C
Minimum detection : N: 0.1 pgN/s, P: 0.03 pgP/s
Dynamic range : 10
3
1.2.5 Auxiliary heated zone
AUX3 to AUX5 : Available (optional)