16 Special Functions
16.2 Batch Schedule
229
GC-2010 Plus
Example 2: For dual AOC
The main tower set up is identical to Example 1.
Sub injector settings are given below.
Vials No. 76 to 78 are analyzed. Chromatopac data processing file No. 0 is set
for the Ch 2.
A standard of known concentration is injected once.
Clean up is performed, but the slope test is not performed.
Sub injector settings Screen display
1. Start vial No.: 76
End vial No.: 78
2. GC file No.: FILE2
3. Data processing file No.:
Ch 2, FILE0
4. Number of injections: 1
Sample type: Standard
5. Slope test: Off
Clean up: On
6. The batch is set up.