Aoralscan 3_Manual_V1.0.0.30 - 14 -
The Aoralscan 3 uses RF energy only for its internal
function. Therefore, its RF emissions are very low and are
not likely to cause any interference in nearby electronic
equipment.
The Aoralscan 3 is suitable for use in all
establishments, including domestic establishments and
those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.
Harmonic emissions IEC
61000-3-2
Voltage fluctuations/flicker
according IEC 61000-3-3
Guidance and manufacturer’s declaration–electromagnetic emissions
Interference immunity
The Aoralscan 3 is intended for use in the electromagnetic environment specified below. The
customer or user of the Aoralscan 3 should assure that it is used in such an environment.
Guidance and Manufacturer's Statement - Electromagnetic emission
Aoralscan 3 is intended to be used in the following electromagnetic environment. The
purchaser or user of Aoralscan 3 should ensure that it is used in this electromagnetic
environment:
Electromagnetic
environment–guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±8 kV contact
±2,±4,±8,±15 kV air
±8 kV contact
±2,±4,±8,±15 kV
air
Floors should be wood,
concrete or ceramic tile. If
floors are covered with
synthetic material, a
relative humidity of at
least 30% is
recommended.
Electrical fast
transient/burst
IEC 61000-4-4
±2 kV for power supply
lines
±2 kV for power supply
lines
Mains power quality should be
that of a typical commercial or
hospital environment.
±0.5, ±1 kV line(s)
to line(s)
±0.5, ±1kV differential
mode
Mains power quality should be
that of a typical commercial or
hospital environment.