www.Siargo.com MF / FS4000 User Manual 20 | P a g e
The detailed information of each register is described below: Y: enabled; N: disabled
Address
0x0001
Write
Y
Read
Y
Description
Address of the product
Value type
UINT 16
Notes
Values from 1 to 255 except for 157 (0x9d).
0 is the broadcast address.
Flow rate
0x0002 ~ 0x00o3
Write
N
Read
Y
Description
Current flow rate
Value type
UINT 16
Notes
Flow rate = [Value (0x0002)*65536 + value (ox0003)]/1000
e.g.: for a flow rate of 123.456 SLPM, the user will read “1 (0x0001)” from
register 0x0002 and “57920 (0xE240)” from register 0x0003, therefore
Current flow rate = (1*65536+57920)/1000 = 123.456
Totalizer
0x0004 ~ 0x0006
Write
Y
Read
Y
Description
Totalizer or accumulated flow rate
Value type
UINT 32 + UINT 16
Notes
A1 = Value (0x0004) * 65536 + Value (0x0005)
A2 = Value (0x0006)
Totalizer or accumulated flow rate = (A1 * 1000 + A2)/1000
e.g.: for a totalizer or accumulated flow rate of 3452.245 m
3
, the user will read
“0 (0x0000)” from register 0x0004; “3452(0x0D7C)” from register 0x0005, and
“245(0x00F5)” from register 0x0006.
Then, the totalizer or accumulated flow rate
= ((0 + 3452)*1000 + 245)/1000=3452.245.
Baud rate
0x0015
Write
Y
Read
Y
Description
Communication baud rate
Value type
UINT 16
Notes
Value=0: 4800; 1: 9600; 2: 19200; 3: 38400. The default value is 3.
Gas conversion factor
0x0016
Write
Y
Read
Y
Description
Gas conversion factor
Value type
UINT 16
Notes
The value is 1000 times the calculated value. The value is used to change the
reference conditions, adjust the system deviation, and calibrations for