www.Siargo.com MF / FS4000 User Manual 21 | P a g e
different gases. It is recommended that write protection should be applied
after this value is changed.
Meter response or data sampling time
The default value is 10 (10 msec). Options are 10, 20, 50, 100, and 200 msec. It
is recommended that write protection should be applied after this value is
changed.
Set the high flow rate alarm limit
The set flow rate value = ((0x0031)*65536+(0x0032))/1000
Set the low flow rate alarm limit
The set flow rate value = ((0x0033)*65536+(0x0034))/1000
Reset or calibrate the offset
UINT 16, fixed value 0xAA55
Send the fixed value 0xAA55 to register 0x0027.
Please make sure there is no flow before executing this command.
Write protection of a specific value
UINT 16, a fixed value of oxAA55
This function disables the write-protected register and allows you to modify
the corresponding register’s value. For example, to change the gas conversion
factor, the user needs to send 0xAA55 to the register 0x0016, and then the
write function will be enabled (write protection is disabled). After the write
execution is completed, the firmware will automatically re-enable the write
protection.