23
If the mattress is to be disconnected from the power supply for an extended period
of time and the mattress is to remain inflated or in the event of a mains power failure,
carry out the following procedure:
• Disconnect from the power unit by pressing the tab on the top of the connector
(1) and pulling away from the control unit (2).
9.8. Mattress Disconnection and Power Cuts
• Switch off the control unit (if still operational).
• Disconnect from the power supply.
Note, there is no need for the air connector to be sealed with a cap due to the use of
the mattress using non-return valve technology.
1. 2.
• The mattress will remain inflated for a maximum of 10 hours only
– Return the system to the mains supply as soon as is practical.
• Whilst unplugged alternating mode will not be operational –
Pressure relief will not be provided.
Warning
Incontinence products such as sheets or pads can be used with the system, however
product performance is likely to reduce due to the reduced effectiveness of the
alternating pressure distribution.
If incontinence products are to be used it is recommended that regular patient skin
checks are performed to ensure skin integrity is maintained.