32
14.1. Requirements According to IEC 60601-2:2007
The Artemis is intended for use in the electromagnetic environment specified below.
The customer or the user of the Artemis should ensure that it is used in such an
environment.
Warning
The Artemis control unit should not be used adjacent to or stacked
with other equipment where possible, if adjacent or stacked use is
necessary the Artemis should be observed to verify normal operation
in the configuration in which it is to be used.
Guidance and manufacture’s declaration – electromagnetic emissions
Emission test Compliance Electromagnetic environment – guidance
RF emission
CISPR 11
Group 1
The Artemis uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
RF emission
CISPR 11
Class B
The Artemis is suitable for use in all establishments,
including domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Harmonic emissions
IEC 61000-3-2
Class A
Voltage fluctuations/
flicker emissions
IEC 61000-3-3
Complies