19
3) Pulsation: Creates tissue stimulus by alternately increasing and decreasing the cell
pressure of the constant low pressure mode by 20% in each direction.
When using the pulsation setting, to return to alternating it is necessary to manually
reselect alternating from the therapy menu, it will not automatically default back.
4) Max inflate: Inflates the cells to maximum pressure (40mmHg) to provide a stable,
static support surface.
The system will automatically revert back to alternation mode after 20 minutes for
patient safety.
When pressed a menu appears offering 3 different settings:
9.5.4.2. Comfort Control
Use the up/down cursor key followed by the selection key to initiate the chosen
setting.
By selecting this function the softness / firmness of the mattress can be manually
altered, dependent on the patient’s requirements. The pump defaults to medium.
Firm = + 5mmHg
Medium = Automated pressure setting
Soft = – 5mmHg
Before changing the pressure a clinical judgement is required from frequent
monitoring and repositioning of the patient.
1: Soft/Firm Control a: Firm
b: Medium
c: Soft
1: Cycle Time
2: Auto Lock
3: Weight Unit (kg or lbs)
Page 1/2
On accessing the comfort control menu ( ) and then pressing the lock ( ) and
down arrow ( ) buttons together a hidden menu is accessed:
9.5.4.3. Additional Settings
4: Weight Display On/Off
5: Egress Alert On/Off
6: Back to the Main Screen
Page 2/3