EasyManua.ls Logo

sidhil Artemis - Specifications

sidhil Artemis
40 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
30
13. SPECIFICATION
Classification: Electrical shock protection: Class I, Type BF
Applied Part: Mattress
Liquid ingress protection: IP21
Not AP or APG equipment*
Supply Rating: 230V, 50Hz, 25W
Fuse Rating:
Mains Plug – 13A
Mains Plug: Type G/BS1363
Mattress Dimensions: (L) 2000mm x (W) 880mm x (D) 200mm
Mattress Weight: 11kg
Maximum Patient Weight: 248kg (39 stone)
No. of Cells: 23 cells which include:
3 static head cells
20 alternating cells (including 8 narrow heel cells) with
cell-on-cell function
Alternating Therapy: AB pattern
Cycle Time: 10, 15 or 20 minutes
Pressure Range: Alternating Mode: 13 – 70 ±2mmHg
Constant Low Pressure Mode: 10 – 40 ±2mmHg
Pulsation Mode: 80% & 120% of CLP setting
Max Inflate: 40 ±2mmHg
Control Unit Dimensions: (L) 180mm x (W) 390mm x (D) 14.5mm
Control Unit Weight: 4.1kg
Cover Material: Dartex®
Cell Material: TPU
Base Material: Nylon fabric with PU coating
Transport and Storage Conditions: Ambient Temp: -25°C to +70°C
Humidity: < 93%, non-condensing
Operational Conditions: Ambient Temp: +5°C to +40°C
Humidity 15% - 93%, non-condensing
Atmospheric Pressure: 700hPa to 1060hPa
Operating Altitude: ≤2000m
Pollution: Degree 2
UV: Intended for indoor use only
Noise level: <40dB(A)
Expected Service Life: 3 years
Safety Standards: IEC 60601-1: 2005
IEC 60601-1-2:2007
IEC 60601-1-11:2010 - Partial
* Not suitable for use in the presence of flammable anaesthetic mixtures with air, oxygen or nitrous oxide.