30
13. SPECIFICATION
Classification: Electrical shock protection: Class I, Type BF
Applied Part: Mattress
Liquid ingress protection: IP21
Not AP or APG equipment*
Supply Rating: 230V, 50Hz, 25W
Fuse Rating:
Mains Plug – 13A
Mains Plug: Type G/BS1363
Mattress Dimensions: (L) 2000mm x (W) 880mm x (D) 200mm
Mattress Weight: 11kg
Maximum Patient Weight: 248kg (39 stone)
No. of Cells: 23 cells which include:
3 static head cells
20 alternating cells (including 8 narrow heel cells) with
cell-on-cell function
Alternating Therapy: AB pattern
Cycle Time: 10, 15 or 20 minutes
Pressure Range: Alternating Mode: 13 – 70 ±2mmHg
Constant Low Pressure Mode: 10 – 40 ±2mmHg
Pulsation Mode: 80% & 120% of CLP setting
Max Inflate: 40 ±2mmHg
Control Unit Dimensions: (L) 180mm x (W) 390mm x (D) 14.5mm
Control Unit Weight: 4.1kg
Cover Material: Dartex®
Cell Material: TPU
Base Material: Nylon fabric with PU coating
Transport and Storage Conditions: Ambient Temp: -25°C to +70°C
Humidity: < 93%, non-condensing
Operational Conditions: Ambient Temp: +5°C to +40°C
Humidity 15% - 93%, non-condensing
Atmospheric Pressure: 700hPa to 1060hPa
Operating Altitude: ≤2000m
Pollution: Degree 2
UV: Intended for indoor use only
Noise level: <40dB(A)
Expected Service Life: 3 years
Safety Standards: IEC 60601-1: 2005
IEC 60601-1-2:2007
IEC 60601-1-11:2010 - Partial
* Not suitable for use in the presence of flammable anaesthetic mixtures with air, oxygen or nitrous oxide.