Symptoms Indications Actions
High Pressure • Audible signal
sounds.
• Screen shows:
Warning!
***HIGH PRESSURE***
1. Ensure the mattress umbilical is not trapped or
being squeezed.
2. Open the mattress and ensure none of the air
pipes are kinked.
3. Turn the unit off and then on again to clear the
indicator.
If a high pressure indicator is still evident turn off at the
mains and contact your approved service provider.
Mattress
Disconnection
• Audible signal
sounds.
• Screen shows:
Warning!
**MATTRESS DISCONNECT**
Please make sure the
connector is securely
fastened
1. Ensure the mattress air connector is correctly
connected to the control unit.
2. Turn the unit off and then on again to clear the
indicator.
If the indicator is still evident turn off at the mains and
contact your approved service provider.
Patient is
bottoming out
1. Ensure the patient is suited to the maximum rating
of the mattress.
2. Ensure the patient is centrally positioned on the
mattress.
3. Increase the pressure setting – Refer to section
9.5.4.2