Installing the ON/OFF assembly 3 - 7
Siemens AG n.a. SP00-000.841.01 Page 7 of 10 ARCADIS
Medical Solutions Rev. 01 08.05 CS PS 24 System Manual
Startup 3
Programming 3
Connect the monitor trolley with the C-arm chassis.
Plug in the power plug.
Switch on the system and wait for the system to boot.
Call up local service at the PC (image processor).
Select "Configuration" and then the Next button to open the configuration window.
Under "Imaging System", select "POWER Off Parameters".
Program the "Power Off Grace Period" parameter to 15 seconds. For this, enter the value
"1500" (1500 x 10 ms corresponds to 15 s).
Select "Save".
Select "OK" to close the "Power Off Parameters were sucessfully saved" window.
Select "Finish".
Select "Home".
Close the local service.
Checks 3
• Switch off the system.
• Switch it on again after approx. 5 to 10 seconds.
- If you press the Off button on the monitor trolley, the C-arm chassis is switched off.
- The PC (image processor) remains switched on.
- If you press the On button on the monitor trolley, the C-arm chassis is switched on
again.
- The PC (image processor) remains switched on.
- After the C-arm chassis has booted, the entire system is ready for operation again.
• Switch the system off with the Off button on the monitor trolley.
• Wait for the PC (image processor) to shut down (after approx. 15 seconds).
• While the PC (image processor) is being shut down, switch the system on again with the
On button.
- The imaging system continues to shut down.
- After that and without the On button being pressed again, the system restarts and is
ready for operation after booting.