E-10 Additional Equipment Manual EEx p
Order No.:
A5E 000 58873-02
3.3 Assembly diagram for the versions EEx p with "Continuous purging"
PE L
N
Power supply
connection
Inlet valve
EEx p purging gas monitoring device
L1
N
PE
Outlet of
purging gas
into Ex-free
zone
Sample
gas inlet
Sample
gas outlet
Inlet of protective gas
(purging gas)
Input pressure: 0,2...0,4 MPa
CALOMAT 6F
OXYMAT 6F
ULTRAMAT 6F
Control electronics
Purging gas
pipelines:
max. 4 m long;
diameter > 8 mm
Gas connections:
C 6F and O 6F: 2 and 4
U 6F (SG): 1 and 2
U 6F (RG): 3 and 4
(see also associated
manual)
5
6*)
7*)
8
Flame inhibitors
Key switch
Choke
Q > 1 l/min
(only with
occasionally
combustible gas
mixtures)
Choke Q > 50 l/min
1) Outlet choke;
dynamic pressure
> 1 hPa
at Q > 1 l/min
2) Overpressure
valve; opens in
the pre-purging
phase
2) 1)
*) see Section 3.1:
"Connection of
sample gas and
protective gas
(purging gas)"