5fMRI User Guide
3 Sequences
3.1 EPI sequences for BOLD Imaging
The following sequences are available as options for acquiring BOLD sensitive images:
ep2d_fid
Basic single-shot EPI sequence
ep2d_bold
Single-shot EPI sequence with the BOLD imaging card paradigm, GLM, t-test, spatial filter, etc.
ep2d_pace
Single-shot EPI sequence with the BOLD imaging card paradigm, GLM, t-test, spatial filter, etc. and prospective
motion correction
Here is a schematic of the EPI data acquisition scheme that is common to all three sequences:
Figure 2. Schematic of data acquisition scheme in EPI sequences. A fat suppression pulse (FS) is followed by a
single-shot acquisition of the acquisition of all the slices. The fat suppression and acquisition of all slices are then
repeated for the number of measurements desired. An optical trigger signal is generated for each measurement.
Optical
trig. out
FS = fat saturation
RF = radiofrequency pulse
PC = phase correction
Single-shot EPI Data acquisition
# of slides
# of measurements (volumes)
Spoil
RF
FS
PC