Page 7—6, Section 7: Troubleshooting
Slides are of variable
thickness
Replace slides
Pale staining Proper stain-to-buffer
ratio has not been
achieved
Refer to Volume and Ratio
Determination on page 6-17
Stain volume is adjusted
too high. Increasing
stain volume may
decrease stain intensity
Refer to Pump Volume
Adjustment on page 6-14
Rinse volume is too high Refer to Volume and Ratio
Determination on page 6-17
Deteriorated or outdated
Stain Pak
Replace Stain Pak