Siemens-Elema AB SPR8-220.092.02 Page 1 of 2 MOBILETT Plus/ -E/ -M/ -HP
Solna, Sweden Rev. 01 08.01 SPS-UD Service
Prerequisites 1
Purpose 1
This document contains additional information on how to remount the collimator on a
MOBILETT Plus/ Plus E/ Plus M/ Plus HP.
Components / documents 1
The remounting changes comprise the following parts:
Tools / documents required 1
Standard installation tools
"Service Instructions", SPR8-220.061.01... or SPR8-215.898/061.01...
Time required 1
Approximately 5 minutes for one customer support engineer
Safety information 1
Part no. Quantity Part no. Name
1 4 60 89 593 Cable ties
2 1 66 08 116 This instruction
Be sure to unplug the unit from the mains or turn off
the main circuit breaker before performing any service.
Remove or install boards only when the generator is
switched OFF. Adhere to the ESD guidelines.
Checks and adjustments performed with radiation ON
are identified by the radiation warning symbol .
During these types of adjustments, radiation protective
clothing must be worn.
Observe all MOBILETT safety guidelines.
CAUTION

Other manuals for Siemens MOBILETT Plus HP

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Siemens MOBILETT Plus HP and is the answer not in the manual?

Siemens MOBILETT Plus HP Specifications

General IconGeneral
BrandSiemens
ModelMOBILETT Plus HP
CategoryMedical Equipment
LanguageEnglish

Related product manuals