MOBILETT Plus/ -E/ -HP SPR8-220.031.04 Page 8 of 30 Siemens AG
Rev. 04 09.03 CS PS 24 Medical Solutions
2 - 8 Installation
10. Place a cable tie (4.7x195mm) around the spiral cable tie from behind (pos. 2 / Fig.
4). Be sure to leave a small loop (pos. 3 / Fig. 8) so they can be positioned in the
lower joint parallel to the two thick cables.
11. Leave the pin plugs for the display (K21) at pos. 4 / Fig. 8.
12. Move the arm to see that the cable runs freely throughout the arm and is not
stretched or trapped. This is especially important in the lower joint.
Collimator assembling 2
1. Dismount the light button and the associated angle bracket (pos. 1 / Fig. 9). Keep
the cable clamp at pos. 2 / Fig. 9 undamaged.
2. Cut the white and grey cables from the “old” light button (pos. 2 / Fig. 9) as close to
the shrinking tubing as possible.
1
Fig. 9 Dismounting the light button and angle bracket