EasyManua.ls Logo

Siemens MOBILETT Plus - Safety Information and Protective Measures

Siemens MOBILETT Plus
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Prerequisites 1 - 3
Siemens AG SPR8-220.031.04 Page 3 of 4 MOBILETT Plus/ -E/ -HP
Medical Solutions Rev. 04 09.03 CS PS 24
Tools / documents required 1
Standard installation tools
Soldering iron
The document Service Instructions for the relevant MOBILETT
Set of ring spanners / open end keys, including a key for 5.5 mm diameter
Drill and drill bit for 5 mm diameter
Time required 1
Approximately 8 hours for one customer support engineer.
Safety information 1
Dangerous x-ray radiation and/or dangerous electrical voltage
during checks and adjustment work steps!
If not observed, death or serious physical injury can occur.
The following items must be complied with when performing ser-
vice work and tests:
- The product-specific safety information in the documents,
- The general safety information contained in ARTD Part 2.
Connect the MOBILETT only to a mains power supply outlet
(receptacle) that corresponds to the installation requirements of
VDE 0107 or to country-specific regulations.
Remove or install boards only when the generator is switched
OFF. Adhere to the ESD guidelines .
Checks and adjustments performed with radiation ON are
identified by the radiation warning symbol . Radiation
protective clothing must be worn for this type of adjustment (see
ARTD, Part 2).
WARNING

Other manuals for Siemens MOBILETT Plus