MOBILETT Plus/ -E/ -HP SPR8-220.031.04 Page 24 of 30 Siemens AG
Rev. 04 09.03 CS PS 24 Medical Solutions
2 - 24 Installation
Installation of power supply, MOBILETT Plus/ -E only! 2
1. For MOBILETT Plus/ -E with a tilt bar (see Fig. 31), remove the tilt bar.
2. If needed, drill two holes of 4.5 mm diameter as shown in Fig. 32. The distances
are: A = 65 mm, B = 40 mm and C = 10 mm. This is only necessary for some older
units.
3. Unplug pin plug K52 from board D8 (pos. 1 / Fig. 33).
4. Fasten the cable shielding connector with the screw holding the protective cover
over board D8 (pos. 4 / Fig. 33). Replace the previously mounted washer with a
washer (KB 5.1mm) included in the kit.
5. Attach pin plug K52b to pin plug K52. Then attach pin plug K52a to K52 on the
board (pos. 1 / Fig. 33).
If drilling is necessary:
Drilling chips can cause injuries to eyes.
Use proper protection.
MOB00644
Fig. 31 Removing tilt bar
MOB00518
A
B
C
Fig. 32 Adjustments for new power supply
CAUTION