Siemens AG SPR8-220.031.04 Page 1 of 4 MOBILETT Plus/ -E/ -HP
Medical Solutions Rev. 04 09.03 CS PS 24
Supplements to Service Instructions 3
Recalibration of the DAP measuring system 3
This instruction shows how to recalibrate the DAP measuring system by using an external
ionization chamber (dose meter).
Required equipment 3
Dose meter
Calibration tool for DAP measuring system, Part No. 65 84 978
Recalibration procedure 3
1. System OFF.
2. Disconnect the DAP ionization chamber connector which is connected to board
D70 (pos. 1 / Fig. 1).
3. Connect the DAP calibration tool between the connector and board D70.
4. System ON.
It is irrelevant which connector is the input or the output of the
DAP calibration tool.
MOB00465
1
Fig. 1 Connector
NOTE