46 Function Check
MOBILETT XP SPR8-230.840.01.05.02 Siemens AG
11.05 CS SD 24
Page 46 of 72
Medical Solutions
6-
5Function Check
Reading out the MOBILETT XP error memory 0
Switch off the system and disconnect the power plug if necessary. Set the XP Hybrid
operating mode switch to "0". Please observe all safety information.
Remove the upper cover. Remove the XP Digital upper covers and place them on their
sides on a table or similar surface. Do not disconnect any cables.
Move switch SW2B on CPU board D916 to position 2 (service) ((Fig. 15 / p. 46)).
Switch on the system.
Service program Pr 1 is displayed on the control panel. Use the mAs +/- keys to select
service program 3. Read out the error memory and make a note of the data (see the
service programs section of these instructions).
Subsequently set switch SW2B of D916 to position 1 (user) (Fig. 15 / p. 46).