EasyManua.ls Logo

Siemens MOBILETT XP ECO - Example: Calculation Radiation Dose

Siemens MOBILETT XP ECO
109 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Radiation Calculation
MOBILETT
SPR8-230.621.01.03.02 65
of 106
Example: Calculation radiation dose
Exposure with U = 70 kV, Q = 20 mAs and source to skin distance, r = 1.05 m:
From Diagram 1, read the corresponding value of Y for U= 70 kV:
Y = 37 µGy x m
2
/mAs
The radiation dose D is calculated with Equation (2):
D = 37 (µGy x m
2
/mAs) x 20 (mAs) / (1.05)
2
=671µGy
The radiation dose at a distance of 1.05 m with tube voltage 70 kV and 20 mAs is
D= 671 µGy.

Table of Contents

Other manuals for Siemens MOBILETT XP ECO

Related product manuals