Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
A5E45779389002-01, 09/2018
19
Figure 4-2 Connections of the CALOMAT 62F analyzer
4.3
Electrical connections
Please observe the following for correct operation:
Provide a disconnector unit in the building installation.
Connect the enclosure of the analyzer to the equipotential bonding (see also associated operating instructions or product
manual).
Use solid or stranded cables with a cross-section of not more than 2.5 mm² for wiring the terminals.
Provide all stranded conductors with end sleeves.
Keep all connection cables within the analyzer short.
Identify the intrinsically-safe lines and route them separately from the non-intrinsically-safe lines. Observe the minimum
distances required when doing this.
Connect the power and signal cables according to the terminal wiring diagrams of the devices to be connected (see the
associated operating instructions or product manuals).
Be particularly careful with the cable inlets (cable glands). Use sealing rings corresponding to the cable diameters for the
cable glands.
Cable diameter ranges and torques M for the screwed glands:
M 20 power cable/RS485: Cable: 7 ... 12 mm; M = 3.8 ± 0.2 Nm
M 20 signal cables: ∅ Cable: 10 ... 14 mm; M = 3.8 ± 0.2 Nm