Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
20 A5E45779389002-01, 09/2018
4.3.1
Connection of the signal cables
NOTICE
Incorrect power supply
The 24 V/1 A power supply must be a power-limited safety extra-low voltage with safe electrical isolation (SELV).
Only connect the signal lines to devices which also have reliable electric isolation from their power supply.
The connection lines to the relay outputs, binary inputs, and analog outputs must be shielded.
The analog outputs are floating, but have a common negative pole.
As a measure to suppress sparking across the relay contacts (e.g. limit relays), RC elements must be connected as
shown in the following figure. Note that the RC element results in a drop-out delay for an inductive component (e.g.
solenoid valve). The RC element should be sized according to the following rule of thumb:
R = R
L
/2; C = 4L/R
2
L
,
where R = 100 Ω and C = 200 nF are sufficient.
You must use a non-polarized capacitor for the RC element.
Figure 4-3 Measure to suppress sparks on a relay contact
When operated with direct current, a spark suppression diode can be installed instead of the RC element.
Refer to the ELAN interface description (Order No. C79000-B5200-C176 German, C79000-B5276-C176 English) for details
on the interface cable.