Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
A5E45779389002-01, 09/2018
41
7.5
Technical specifications CALOMAT 62F
Design, enclosure
Weight (analyzer)
Approx. 25 kg (55 lbs.)
Electrical characteristics
Power consumption
For non-heated gas connection block
For heated gas connection block
Approx. 30 VA
approx. 330 VA
Fuse values
For non-heated gas connection
100 ... 120 V
200 ... 240 V
For heated gas connection
100 ... 120 V
200 ... 240 V
T1L250V (F3/F4)
T630L250V (F3/F4)
T1L250V (F1); T4L250V (F2/F3/F4)
T630L250V (F1); T250L250V (F2/F3/F4)
Gas path
Gas connections
Female thread 1/8"-27 NPT (according to ANSI B 1.20.1)
Gaskets (O-rings)
FPM (e.g. Viton) or FFPM (e.g. Kalrez)
Sensor
Glass
Internal gas path (gas lines)
Stainless steel 1.4571 or Hastelloy C22
Leak-tightness
Loss through leaks < 0.5 μl/s
Gas inlet conditions
Permissible sample gas pressure
800 to 1100 hPa (11.6 to 16 psi) absolute
Sample gas flow
30 to 90 l/h (0.5 to 1.5 l/min)
Sample gas temperature
For non-heated gas connection
For heated gas connection
0 to 50 °C (32 to 122 °F)
0 to 80 °C (32 to 176 °F)
Temperature of measuring cell
60 °C (140 °F)
Temperature of gas connection with heated gas
inlet
Approx. 70 °C
Sample gas humidity
< 90% RH (relative humidity), non-condensing
Purging gas pressure < 165 hPa