Compact Operating Instructions in accordance with EN 61010-1 and EN 60079-0
44 A5E45779389002-01, 09/2018
Sample gas inlet conditions
Sample gas inlet conditions of analyzer versions
Type of gas Non-combustible gases or gas mixtures that
always remain below the LEL in accordance
with Category 4 from EN ISO 13849-1: 2016
Combustible gases up to gas mixtures which
are occasionally explosive
Max. pressure p
max
1100 hPa for CALOMAT 6F and
CALOMAT 62F
3000 hPa for OXYMAT 6F
1500 hPa for ULTRAMAT 6F
For combustible gases and analyzer ver-
sions Ex px with "leakage compensation"
100 hPa (rel.) for ULTRAMAT 6F and
OXYMAT 6F
Analyzer versions Ex px with "continuous
purging" as well as
1100 hPa for CALOMAT 6F and
CALOMAT 62F
3000 hPa for OXYMAT 6F
1500 hPa for ULTRAMAT 6F
Min. pressure p
min
500 hPa for OXYMAT 6F
600 hPa for ULTRAMAT 6F
800 hPa for CALOMAT 6F and
CALOMAT 62F
For combustible gases and analyzer ver-
sions Ex px with "leakage compensation"
-200 hPa (rel.) for ULTRAMAT 6F and
OXYMAT 6F
Temperature CALOMAT 6F: 0 to 50 °C
CALOMAT 62F: 0 to 50 °C
OXYMAT 6F: 0 to 50 °C
OXYMAT 6F: In heated version: 0 to 130 °C, but only 15 °C above temperature analyzer unit
ULTRAMAT 6F: 0 to 50 °C
ULTRAMAT 6F: In heated version: 0 to 80 °C
Sample gas flow rate q
v
CALOMAT 6F / CALOMAT 62F: 0.5 ... 1.5 l/min
OXYMAT 6F: 0.3 ... 1.0 l/min
ULTRAMAT 6F: 0.3 ... 1.5 l/min