EasyManua.ls Logo

Siemens SD 270 - Insert Masking Earpiece Calibration Method; Transducer Types and Reference Levels; Environmental and I;O Specifications

Siemens SD 270
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
41
Insert Masking Earpiece
Calibration method: With 2cc coupler compliant with
IEC 126
Transducers
Types and reference levels: DD45: ISO 389-1, Table 2
E-5A: ISO 389-2, Table 1
B-71: ISO 389-3, Table 1
Static headband force: Headphones: 4.5N
Bone vibrator: 5.4N
Bone vibrator calibrated: For mastoid placement
Sound attenuation
characteristics: ISO8253-1, Table 3
Airborne sound from bone
vibrator: See Br. J. Audiol. 1980, P73-75
Environmental
Operating temperature: +15°C to +35°C
Operating humidity: 30% to 90% (non-condensing)
Atmospheric pressure: 700 hPa to 1060 hPa
Input / Output
Power input: 2.5mm barrel-type socket
Mic inputs impedance: 2500 Ohms
Mic inputs connection: Mono 3.5mm Jack socket
Line input impedance: 6800 Ohms
Line in/out connection: Stereo 3.5mm Jack sockets