EasyManua.ls Logo

Siemens SD 270 - Live Voice Speech Audiometry with Masking

Siemens SD 270
84 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
53
2. Live Voice Speech Audiometry with Contralateral Masking
2.1 Set Up ‒ as described in 1.1 then
f Select MASK
f Speech-weighted masking is now routed to the opposite
earphone to that selected.
f The MASK rotary control changes the masking level in 5dB steps.
f If required, readjustment of the input signal level can be
accessed by temporarily deselecting the MASK key ‒ then
proceed as in 1.1d).
2.2 Procedure
As described in 1.2 but adjusting the masking level as required.

Table of Contents

Related product manuals