SIREMOBIL Compact Register 3 RXR2-130.033.01 Page 4 of 10 Siemens AG
Rev. 04 04.02 CS PS 24 Medical Engineering
4 Installation and Setting Instructions
Installation 1
Removing the covers 1
• Disconnect the SIREMOBIL line voltage plug.
• Remove the three cover plates from the basic unit.
• Remove the SIREPHOS cover.
First, carefully remove the adhesive strip on the side panels of the cover.
Installing the area dose measurement chamber 1
• Pull the measurement chamber connector (1/Fig. 1) out of the cable tie.
• Place the area dose measurement chamber on the SIREPHOS as shown in Fig. 1.
The "Focus" label must point toward the X-ray tube.
• Secure the area dose measurement chamber to the SIREPHOS with the enclosed
screws.
• Plug the measurement chamber connector (1/Fig. 1) into the measurement chamber.
Installing the measurement board D16 1
• Install measurement board D16 according to Fig. 2.
• Plug the two connectors (2/Fig. 2) into PC board D16.
• Connect the enclosed ribbon cable to D1.X9 and D16 (3/Fig. 2).
Fig. 1 Fig. 2
2
2
3
1