SIREMOBIL Compact Register 3 RXR2-130.033.01 Page 10 of 10 Siemens AG
Rev. 04 04.02 CS PS 24 Medical Engineering
10 Installation and Setting Instructions
Final Steps 1
• Switch off the SIREMOBIL.
• Reinstall all covers.
• Perform the protective conductor test according to ARTD - 022.731.17.. .The protective
conductor resistance must not exceed 0.2 Ohm.
Changes to previous version 1
Page 2 Measuring devices actualized.
Page 6 Corrections in descriptions added.
Page 7 Corrections in descriptions added.
Page 9 Maximum values for the protective conductor resistance added.