Installation and Setting Instructions 9
Siemens AG Register 3 RXR2-130.033.01 Page 9 of 10 SIREMOBIL Compact
Medical Engineering Rev. 04 04.02 CS PS 24
Evaluation 1
Calculate the deviation between the area dose product measured and the area dose
product displayed:
The deviation must be < 0.3.
Deviation = [(displayed area dose product FDPa) - (measured area dose product FDPg) ]
/ (measured area dose product FDPg).
For deviations > 0.3, replace the area dose measurement device.
Fig. 3 Fig. 4
Fig. 5
Meas. chamber
Sirephos
I.I.
Centering cross