SIREMOBIL Compact Register 3 RXR2-130.033.01 Page 6 of 10 Siemens AG
Rev. 04 04.02 CS PS 24 Medical Engineering
6 Installation and Setting Instructions
Programming and adjusting the area dose measurement device 1
Calculating the CONSTANT parameter (Qt * Kg) 1
• Read and record the Kg value shown on the electronics unit of the area dose
measurement device.
• Read the Qt value on the area dose chamber of the area dose measurement device.
• Multiply these values and record them. The resulting value is the CONSTANT (QT*Kg)
value to be programmed.
Programming 1
• Connect the service PC to the serial interface on board D1.
• Start the service program.
• Select the Adjustment, Parameter menu.
• Select "Diamentor" in the combo box "Parameter Groups:".
• Click "Get from Unit".
• Select "1. Diamentor Existing".
• Select ’yes’ in the "Value actual" field.
• Click "Set Value".
• Select "2. Constant (Qt * Kg)" .
• Enter the parameter you have calculated in the "Value Actual:" field.
• Click "Set Value".
• Click "Put to Unit".
• Click "Save to File".
This adjustment is used to calibrate the area dose measurement
device of the SIREMOBIL Compact display.