SIREMOBIL Compact Register 3 RXR2-130.033.01 Page 6 of 10 Siemens AG
Rev. 04 04.02 CS PS 24 Medical Engineering
6 Installation and Setting Instructions
Programming and adjusting the area dose measurement device 1
Calculating the CONSTANT parameter (Qt * Kg) 1
Read and record the Kg value shown on the electronics unit of the area dose
measurement device.
Read the Qt value on the area dose chamber of the area dose measurement device.
Multiply these values and record them. The resulting value is the CONSTANT (QT*Kg)
value to be programmed.
Programming 1
Connect the service PC to the serial interface on board D1.
Start the service program.
Select the Adjustment, Parameter menu.
Select "Diamentor" in the combo box "Parameter Groups:".
Click "Get from Unit".
Select "1. Diamentor Existing".
Select ’yes’ in the "Value actual" field.
Click "Set Value".
Select "2. Constant (Qt * Kg)" .
Enter the parameter you have calculated in the "Value Actual:" field.
Click "Set Value".
Click "Put to Unit".
Click "Save to File".
This adjustment is used to calibrate the area dose measurement
device of the SIREMOBIL Compact display.