DEVICE DESCRIPTION
2
15
SITRANS FX330
www.siemens.com/flow09/2018 - A5E40861058-AB
Max. permissible product and ambient temperatures with signal converter in stainless
steel (bright) or connection box in stainless steel (bright) mounted above the flow
sensor
Temperature class in °C
Temperature class in °F
Temperature class T6 T5 T4 T3 T2
T
amb
in °C 60 65 60 65 60 65 40 60 65 40 60 65
Nominal size
DN15…25 75 65 100 100 135 135
1
200 170
1
145
1
225 170
1
145
1
DN40…50 75 65 100 100
1
135
1
130
1
200 150
1
130
1
235 150
1
130
1
DN65…100 75 65 100 100
1
135
1
115
1
195
1
135
1
115
1
195
1
135
1
115
1
DN150...300 75 65 100 100
1
135
1
135
1
200 155
1
135
1
230 155
1
135
1
1 Permanent service temperature of connecting cable and cable entry min. 80°C
Temperature class T6 T5 T4 T3 T2
T
amb
in °F 140 149 140 149 140 149 104 140 149 104 140 149
Nominal size
DN15…25 167 149 212 212 275 275
1
392 338
1
293
1
437 338
1
293
1
DN40…50 167 149 212 212
1
275
1
266
1
392 302
1
266
1
455 302
1
266
1
DN65…100 167 149 212 212
1
274
1
239
1
383
1
274
1
239
1
383
1
274
1
239
1
DN150...300 167 149 212 212
1
275
1
275
1
392 311
1
275
1
446 311
1
275
1
1 Permanent service temperature of connecting cable and cable entry min. 176°F