What to do if Siemens Measuring Instruments display shows no flow?
- KkarlwilliamsAug 2, 2025
If the Siemens Measuring Instruments display shows no flow, it could be due to several reasons: the pipe is not filled with liquid (ensure the pipe is filled, or at least the sound path is covered with liquid); a broken or detached cable (ensure the sensor cable is connected in the terminal housing of the transmitter and that the screen and center pin are connected inside the sensor top); or a permanent disturbance inside the pipe (if possible, ensure that the sensor window inside the pipe is not damaged and that no permanent disturbance is present).