SIMATIC PDM
B.1 Commissioning with SIMATIC PDM
SITRANS F Ultrasonic flowmeters FST030 (HART)
134 Operating Instructions, 08/2017, A5E35971532-AC
Table B- 1 Device menus
Shows the communication interface (HART modem)
Communication (Modbus) Shows the communication interface (Modbus interface)
Downloads all writable parameters to the device
Upload to PC/PG Uploads all parameters from the device to the parameter
table
Update Diagnostic Status Reads current diagnostic status from the device and
updates the diagnostic status icon
Sets the HART polling address
Guide for a quick commissioning
Wizard - Clamp-On Configuration...
Guide for a configuration of clamp-on sensors
Wizard - Inline Configuration...
Guide for a configuration of inline sensors
Trim Signal Output (online dialog)
Calibration of current output (channels 2 to 4)
Totalizer (online dialog)
Controlling totalizers 1, 2 and 3
Multipoint calibration The multipoint calibration allows the user to set a calibra-
tion curve, if deemed necessary, at various flow rates
across the range of operation
FW update A FW update is a firmware update. This function allows
the user to install firmware that has been placed on the
Maintenance (online dialog)
Setup of maintenance functions
Suppress Alarms Suppresses individual alarms
Audit Trail Lists parameter changes, FW updates, and alarm history
Simulation (online dialog) Simulation of process values, alarms, and inputs/outputs
(channels 2 to 4)
Access Management Possibility to upgrade access level from User to Expert
and to change PIN code for "expert" level
Resets device to default settings or restarts device
Table B- 2 View menus
Process Variables (online dialog)
Device Diagnostic (online dialog) Shows all diagnostics information (alarms and advanced
Advanced diagnostics Advanced diagnostics provides the user with operational
information to assist in determining that the device is
functioning properly, producing quality data, or assist in