Applications
A.1 Liquids applications
Transducers
44 Compact Operating Instructions, 09/2021, A5E43390688-AC
A.1.4 Volume
Main fill level for full calibration. Above this
level erroneous readings will result as level has
entered blanking zone.
Span: Corresponds to tank manufactur-
er's empty level
'Alternate'
Rise
'Preferred'
Tank manufacturer's empty level
'Bad'
Discharge
Tank manufacturer's full level May require target to obtain empty
reading
Beam angle
Empty level for 'alternate' locations.
Below this level, echo would reflect
away from the transducer.