Commissioning
6.5 Local commissioning
SITRANS LT500 with ultrasonic transducer inputs
98 Compact Operating Instructions, 02/2024, A5E52021877-AA
Step: Alarms for relay 1
Step is repeated per selected relay
This parameter must be set for each relay selected in "Step: Alarm relays".
Only the parameter for relay 1 is shown here as an example.
Alarms for relay 1
Enables selected alarms per relay.
List of available alarms is based on selections in "Step: Enable alarms", and if "Sensor
diagnostics" was enabled in "Step: Alarm sources".
All alarms are disabled by default, and as each alarm is enabled for the current relay, it is no
longer available for the next relay in this step.
Selections made here, including those made to a relay already configured for another
application, will take effect when the wizard is completed.
Step: Apply?
Apply?
Applies settings as last step in wizard.
Wizard completes and settings are applied.
Returned to start of wizard.