EasyManua.ls Logo

Siemens SITRANS P series - Construction Details - Pressure Types; Construction - Flush-Mounted Diaphragm; Construction - PMC Connection

Siemens SITRANS P series
288 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SITRANS P, DS III series (7MF4.33.. 7MF4.34.. 7MF4.35..)
A5E03434631-02, 12/2012
41
Rated conditions DS III with PMC connection
Installation conditions
Ambient temperature
Note Observe the temperature class in hazardous areas.
Measuring cell with silicon oil
filling
-40 … +85 °C (-40 … +185 °F)
Display
-30 … +85 °C (-22 … +185 °F)
Storage temperature
-50 … +85 °C (-58 … +185 °F)
Climate class
Condensation Permitted
Degree of protection in
accordance with EN 60529
IP65, IP68
Degree of protection in
accordance with NEMA 250
NEMA 4X
Electromagnetic compatibility
Emitted interference and
interference immunity
In accordance with EN 61326 and NAMUR NE 21
Process medium conditions
Process medium temperature
-40 … +100 °C (-40 … +212 °F)
Rated conditions for absolute pressure (from the differential pressure series), differential pressure and flow rate
Installation conditions
Installation instruction
any
Ambient conditions
Ambient temperature
Note Observe the temperature class in hazardous areas.
Measuring cell with silicon oil
filling
-40 … +85 °C (-40 … +185 °F)
Measuring cell
30 bar (435 psi)
-40 … +85°C (-40 … +185 °F)
for flow rate: -20 … +85°C (-4 … +185 °F)
Measuring cell with inert liquid
filling
-20 … +85 °C (-4 … +185 °F)
Display -30 … +85 °C (-22 … +185 °F)
Storage temperature -50 … +85 °C (-58 … +185 °F)
Climate class
Condensation Permitted
Degree of protection in
accordance with EN 60529
IP65, IP68
Degree of protection in
accordance with NEMA 250
NEMA 4X
Electromagnetic compatibility
Emitted interference and
interference immunity
In accordance with EN 61326 and NAMUR NE 21
Process medium conditions
Process medium temperature
Measuring cell with silicon oil
filling
-40 … +100 °C (-40 … +212 °F)
Measuring cell
30 bar (435 psi)
-40 … +85 °C (-40 … +185 °F)
With flow: -20 … +85 °C (-4 … +185 °F)