9.2.33.1 Reset to sensor calibration
Resets the zero point and sensor calibration to the factory setting.
9.2.33.2 Reset DAC trim to the factory setting
Resets the DAC trim (digital-to-analog converter trim) to the factory setting.
The
DAC trim is used to calibrate the 4 mA and 20 mA end points of the analog output with an
external reference (e.g. current measurement device).
The DAC trim is available over remote operation.
9.2.33.3 Restore ordered configuration
With this function you return your device to its delivery state.
The ordered configuration of the following parameters is restored:
Pressure units
Quick start
Pressure reference
Long tag (TAG)
Short tag (TAG)
Lower range value
Upper range value
Lower limit of the measuring range
Upper limit of the measuring range
Damping value
Application and related values (e.g.: vessel shapes and application points)
Fault current selection
Lower fault current
Upper fault current
Lower saturation limit
Upper saturation limit
Custom units
The parameters that you have not configured via the order are reset to the factory settings.
9.2.33.4 Restore factory settings
With this function you return your device to its factory setting.